Measurement & Analysis




 

 
Model Specification
Internal Dim. W x H x D (cm) 50x60x50
External Dim. W x H x D (cm) 95x150x105
* Temp. & Humi. Range -40C~ +150C & 10RH% ~98% RH
 
Accelerated aging of sterile barrier systems for medical devices(Standard Guide: ASTM F1980-16)
Conversion table for temperature vs testing time
Temp( C) Accelerated aging time
(Week)
Equivalent real time aging
(Year)
(1)Ambient temperature:25C
(2)Q10=2.(Conservative aging factor)
55C 6.5   Week 1 Year
55C 19.5 Week 3 Year
55C 32.5 Week 5 Year
*Temperature range 50,55或60C
 
Environmental test (Standard Guide: IEC 60068-2. GB/T 2423)
Items standard Object
Cold test IEC 60068-2-1
GB/T 2423.1
Testing for electric and electronic products operation capacity under cold environment.
Dry heat test IEC 60068-2-2
GB/T2423.2
Testing for electric and electronic products operation capacity under heat environment.
Damp heat, steady state test IEC 60068-2-78
GB/T 2423.3
Testing for electric and electronic products operation capacity under damp heat and steady state environment.
Composite temperature
/ Humidity cyclic test
IEC 60068-2-38
GB/T 2423.34
Testing for electric and electronic products operation capacity under composite temperature /humidity cyclic environment.
Storage product temperature cycle test and temperature range: